DIMENSIONS
Dimensions
PANEL CUTOUT
Figure 8-1. Panel Cut-Out

SIGNAL INPUT
Input Ranges
0-100 mV, ▒ 50 mV, 0-10 V, +5 V, 0-20 mA, 4-20 mA

Isolation to Earth Ground
  • 354 V peak per IEC spacing
  • (3mm) 1500 Vac/dc minimum per 1 min. High Voltage test


  • Noise Rejection

  • Normal Mode Rejection (NMR) = 60dB
  • Common Mode Rejection (CMR) = 120dB


  • Protection
  • 240 V rms max for voltage input ranges
  • 200 mA for current input ranges


  • Resistance

  • 100 Meg ohms for 100 mV or +50 mV range
  • 1 Meg ohm for 10 V or +5 V range
  • 5 ohms for 20 mA current input


  • Display
    LED 14-segment, 13.8 mm (0.54")

    Symbols: 8. 8. 8. 8.

    ANALOG TO DIGITAL
    Technique: Dual slope
    Internal resolution: 15 bits
    Read Rate: 3/sec Polarity Automatic

    ACCURACY AT 25░C
    Max Error Strain/Process: ▒0.03% of reading, ▒ 1 count
    Span Tempco: 50 ppm/░C
    Step Response: 1 sec
    Warm Up to Rated Accuracy: 30 min

    Excitation Voltage
    AC power units
    DC power units
    24 V @ 25 mA
    24 V @ 25 mA
    12 V @ 50 mA
    12 V @ 50 mA
    10 V @ 120 mA
    10 V @ 60 mA
    5 V @ 60 mA
    5 V @ 60 mA

    Load Regulation: 1.1%
    Line Regulation: 0.02% per Vac


    ALARM OUTPUTS (if applicable)
    2 Form "C" on/off relays. Configurable for latched and unlatched by software.
  • Max current: 5 AMPS, resistive load
  • Max voltage: 250 Vac or 30 Vd

  • ANALOG OUTPUT (if applicable)
    Signal Type: Current or voltage

    Signal Level

  • Current: 10 V max compliance at 20 mA output
  • Voltage: 20 mA max for 0-10 V output


  • Function
    May be assigned to a display range or proportional control output with setpoint #1 when used as a control output.

    Linearity: 0.2%

    Step Response Time

    2 - 3 seconds to 99% of the final value

    ISOLATED ANALOG OUTPUT (TB5, if applicable)
    Same as non-isolated analog output except isolated to 1000 Vdc.

    Signal Type: Current or voltage

    Signal Level

  • Current: 10 V max compliance at 20 mA output
  • Voltage: 20 mA max for 0-10 V output


  • Function
    May be assigned to a display range or proportional control output with setpoint #1 when used as a control output.

    Linearity: 0.2%

    Step Response Time

    2 - 3 seconds to 99% of the final value

    NOTE: Only one analog output is available on each unit and it must be factory installed.


    INPUT POWER INFORMATION
    Voltage
  • AC - 115/230 Vrms ▒10%
  • DC - 9.5 at 950mA dc to 32 V dc at 250mA dc


  • Frequency: 50-60 Hz
    Power: 9.5 watts max.

    External Fuse Protection Recommended
    AC
    115V
    IEC 127-2/III, 125mA, 250V (T Type) or UL Slow-Blow, 175mA, 250V
    AC
    230V
    IEC 127-2/III, 63mA, 250V (T Type) or UL Slow-Blow, 80mA, 250V

    ENVIRONMENT
    Operating Temperature: 0░ to 50░C (32░ to 122░F)
    Storage Temperature: -40░ to 85░C (-40░ to 185░ F)
    Relative Humidity: 90% at 40░C (non-condensing)

    MECHANICAL
    Panel cutout: 1/8 DIN 3.62" x 1.78" (45 x 92mm)
    Weight: 1.27 lb (575 g)
    Case material: Polycarbonate, 94 V-O UL rated

      Help Price List Specifications Print Manual Manuals Home
     Home Home
     Table of Contents TOC  Back Back  Next Next